Medela Manual Base Pump

$75.00

  • Easy to use
  • Individually adjustable vacuum
  • Portable and lightweight
  • Single pumping

or 6 weekly interest-free payments from $12.50 with Laybuy what's this?

Found this product at a lower price?

Description

Key Features
  • Operation with physiologically natural suction principle-Vacuum release after each piston pulling.
  • Can be upgraded to a Mini Electric.
  • City Style Breast pump Bag available.
Recommendation for use
  • Ideal for short-term use.
  • 2-3 times per week.
  • Single user product-personal care item.
Technical Data
  • Suction strength Full range vacuum adjustable.
  • min. approx-100 mmHg.
  • max. approx-240 mmHg.
  • Piston/Cylinder System.
Breastpump care
  • Unplug breastpump prior to cleaning.
  • At home Daily or when spills occur.
  • In between each mother and/or on a regular basis, e.g. daily.
  • Use a damp (not wet) cloth to wipe over the breastpump. Optionally, soapy water or a non-abrasive detergent may be used.
  • Do not spray or pour liquid directly onto the pump.
  • Do not immerse the Symphony pump unit in water or shower the diaphragm area.
Personal Hygiene
  • Hands should be thoroughly washed with soap and water before touching the breast and pump parts.
  • The breast and surrounding area needs cleaning with a damp cloth. Soap or alcohol should not be used which can dry the skin and cause cracked nipples.
  • Touching the inside of bottles and lids should be avoided.
Useful Hints/Trouble Shooting
  • Always start pumping at minimum level.
  • If there is no vacuum or if the pumping performance is poor, make sure the pumping system is closed (no leakages).
  • The rubber seal needs to be intact and slipped correctly onto the piston; a damaged rubber seal causes less vacuum and needs replacing.
  • Make sure the cylinder is tightly screwed on the cylinder holder.
  • The valve membrane needs to be clean and fit properly on the valve head; a damaged valve membrane causes less or no vacuum and needs replacing or proper cleaning.
  • Check proper fit of valve head on breast shield.
  • Check proper fit of breast shield on breast (needs to seal completely).
  • The cylinder has to be clean on the inside, otherwise the rubber seal can be damaged.